Technical Data - Beurer SL 30 Instrucciones Para El Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 14
Important information – please keep for later use
• Read these notes carefully.
• Caution – if operating equipment or procedures are used which differ from those specified here, this can result in dangerous exposure to radiation.
• Never look into the laser beam, not even with optical instruments (e.g. glasses, contact lenses, magnifying glass, microscope etc.) from a distance of 100 mm or less.
If this happens, immediately consult an eye specialist. The unit is a Class 2M laser.
• Do not use on small children or vulnerable individuals.
• Patients who have previously had a face lift must consult their skin specialist before using the Softlaser.
• If you are uncertain about its use, we recommend you consult your doctor.
• The laser beam must not be directed against other persons.
• The laser unit must be stored out of the reach of children.
• The laser unit must be secured against improper use by unauthorised persons.
• The unit should not be used with anaesthetic gases or flammable substances.
• The batteries should be removed if the unit is not to be used for some time. Replace batteries once the unit has been used approximately 100 times.
• The visible metal parts are manufactured from medical-grade steel. If used and cleaned correctly no allergic reactions are therefore to be expected.
• For reasons of hygiene, the softlaser should only be used on one person.
• Repairs may only be carried out by the specified authorised repair centres. Caution: Incorrect handling and faulty repairs may lead to a dangerous exposure to radiation.
• The laser unit must not be used in clinical areas or in the vicinity of strong electromagnetic fields (e.g., close to radiation units or mobile telephones etc.).
Technical Data
Unit class:
Laser class 2M
IEC 60825-1: 1993 + A1: 1997 + A2: 2001
Laser:
GaAlAs (Gallium-Aluminium-Arsenide) diode, continuous beam
Power output:
max. 5 mW
Laser wavelength:
635 – 670 nm
Power supply:
2 x 1.5 V alkaline batteries ("micro": AAA / IEC Type LR 03)
Operating environm.: 10 °C to 40 °C at a relative humidity of 30% to 90%
Storage environment: -10 °C to + 60 °C at a relative humidity of 35% to 75%
The device does not belong to the category AP/APG.
Should additional technical questions arise, please contact: Beurer GmbH & Co. KG, Söflinger Straße 218, 89077 Ulm, Germany
Rated voltage:
Operating mode:
Class of protection:
Explanation of symbols:
10
3V DC
Continuous operation
internal supply, IPXO
Read Instructions for Use!
Degree of protection BF

Publicidad

Tabla de contenido
loading

Tabla de contenido