2. Using a suitable crosshead screwdriver, undo the 8
screws in the back panel.
3. Place the body so that the front ofthe glass ceramic
plate is lying on the flat surface.
4. Take the cool lamp by the left or right end and press
it carefully to the right or left.
5. Now lift it forwards and out of the socket.
NOTE
Use a cloth to hold the new lamp when inserting
it, not bare fingers, as the grease on your skin
could damage the coating on the lamp.
6. Insert the new lamp in the socket. The new lamp is
inserted in the reverse order to removal.
7. Close the unit again. The unit is closed in the reverse
order to opening it.
Make sure that no foreign objects are left in the
body of the unit and that nocables become trapped.
Replacement lamp
Type: Halogen lamp 110 V – 130 V / 300 W
118 mm long for R7S socket
Original Order no. 162.783
The service life of the lamp should be approximately
2000 hours

9. Storage

Remove the plug from the wall outlet and let the
unit cool down completely before you move it or
pack it away.
If possible, store the infrared heat lamp in its original
packaging in a dry place.

10. Disposal

Dispose of the device, components and optional ac-
cessories according to applicable local regulations.
Unlawful disposal may cause environmental pollu-
tion.
In the interests of environmental protection the infrared
lamp should not be discarded with household garbage
at the end of its service life. Dispose of the unit at a
suitable local collection or recycling point.
Follow the local regulations for disposal of materials.
Dispose of the unit in accordance with the local Di-
rective for waste electrical and electronic equipment.
If you have any questions, please contact the local au-
thorities responsible for waste disposal.
This equipment complies with European and US Stan-
dards EN 60601-1-2 and is subject to specific safety
provisions for electromagnetic compatibility. Please
note, however, that portable and mobile HF commu-
nication equipment may influence this unit. Further de-
tails may be obtained from our Customer Service De-
partment.

11. Technical specifications

Voltage
AC 110 V – 120 V / 60 Hz
Power
300 W
consumption
Timer
Adjustable from 1 to 15 minutes in
one-minute steps.
Dimensions
10.6 x 11.2 x 7.7 in
(W x H x D)
(270 x 285 x 195 mm)
Weight
Approx. 3.5 lbs (1.6 kg)
Operating
Temperature: 50 – 95°F
conditions
(10°C to 35°C)
Relative humidity: ≤ 90%
Storage and
Temperature: 14 – 122°F
transport
(-10°C to 50°C)
Relative humidity: ≤ 90%
Protection class II
Radiation
max. radiation intensity at a dis-
intensity
tance of 11.8 in (30 cm)
490 W/sq m
IR type
A + B