Use ONLY a Heathrow Scientific
2.
This rotor must be installed in the unit for this
function to work. Attempted use of any other
rotors will void the warranty and may cause
damage to the device or the surroundings.
Use of any other rotors may cause
INJURY. This rotor has two silver metal
magnets visible on the top surface:
3.
If the wrong rotor is installed:
a. If you wish to use the stirring function and the magnetic stirrer
rotor is not installed. The unit will detect that the incorrect rotor is
installed within 2 seconds after the rotor starts spinning. A tone
will sound and the rotor will stop rotating. The display will show
"ERR: Rotor Type".
b. After removing any vessel or other object resting on the unit; the
lid can be opened by pressing and holding the control knob for 2
seconds to actuate the lid release (you will hear a click or clicking
sound indicating the latch has been released) and the lid will open
slightly.
c. Tilt the lid upwards and to the rear of the device.
d. Loosen the retaining knob in the middle of the rotor and remove
the incorrect rotor.
e. Install the magnetic stirrer rotor. Make sure to tighten the knob in
the center of the rotor.
f. Close the lid. Make sure it latches.
4.
Make sure power is off on the main power switch on the rear of the
housing.
5.
Plug the power adapter plug into the connector on the rear of the
MagFuge and make sure the green light on the power adapter is
illuminated.
6.
Turn on the main power switch
7.
The display will immediately light up to display the MagFuge name
and the Software/firmware version followed by the message "Press
to start". If the magnet stirring rotor is not installed in the unit, go to
step 3.
8.
Press the control knob and release. The message will change to
"Mode" followed by "Centrifuge."
9.
Rotate the control knob to select "Stirrer".
10. Press the control knob and release. The message will change to "Mix
Direction:" followed by "Clockwise"P.
11. If you wish to select clockwise rotation P, go to step 12. To change
the rotation direction to counter clockwise Q, rotate the control knob
to change the message to "C. Clockwise" (counter clockwise Q).
12. Press and release the knob to change the display to "Speed".
Note: Speed can be adjusted while the unit is running.
14
magnet stirrer rotor when stirring.
®