19. To cancel rotation. Press the control knob for 2 seconds and release.
A tone will sound and the rotor will immediately slow to a stop.
20. When the time has counted down to 00':00" a tone will sound and
the rotor will slow to a stop. To prevent spillage of the stirring vessel
on the lid, the lid will remain latched.

Error Messages:

If an error happens, the unit sound a tone beep and the display will
indicate the error.
Error Status
Motor Overload
Balance (when
centrifuging)
Temperature
Excessive Tilt
Cover Open
Rotor Type
Resolution
— Something is interfering with the rotor. Clear
the rotor and reset.
— Inspect the tubes for equal tube fill or
improper placement. Once corrected, rerun.
— If the balance error continues to happen,
remove the tubes and determine if the
balance error still persists with an empty
rotor.
— Inspect the rotor for improper installation.
— The unit has exceeded the normal
operating temperature.
— Turn off the unit and allow to cool.
— The unit has experienced a non-normal tilt
event.
— Make sure the unit is placed on a level
surface. Once corrected, rerun
— The lid has opened during the cycle.
Check for proper operation of the lid latch
mechanism. The lid should stay locked
during the entire cycle.
— The wrong rotor is installed for the selected
mode (ex. Centrifuge rotor installed when
stirring or oscillating modes are selected).
— Press and hold the control knob for 3
seconds to release the lid latch (a click or
clicks will be heard). Open the cover and
install the correct rotor.
Rotors are retained by a central threaded
knob. Grasp the knob between your fingers
and turn the knob counter clockwise Q to
unscrew the knob.
Tighten the knob by turning it clockwise P
until an increase in resistance is felt.