8.5. Capacitance measurement
The capacitance value is obtained by measuring the insulation resistance.
After finishing measuring (When the key has been pressed), wait a
few seconds and then press the
exhibited on the display.
Voltage
500 V
1,000 V
2,500 V
5,000 V
10,000 V
Precision: ± 10% of the measured value ± 3 digits
Note: "0" will be exhibited on the display when measuring values lower
than 50 nF
8.6. Hold
This function allows holding the last performed reading on the display at
the moment when pressing the
When this key is pressed again, the equipment updates the resistance
and chronometer values. The led of
display indicate that the function has been activated
key. The capacitance value will be
Capacitance
50 nF up to 10 µF
50 nF up to 5 µF
30 nF up to 2 µF
30 nF up to 1 µF
30 nF up to 680 nF
key, without interrupting the test.
key and the letter H on the
28