NSK Ti Max nanoSG20LS Operation Manual página 10

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

7-1 Cleaning (Handpiece)
1) Remove the bur/drill form the handpiece.
2) Half fill a cup with clean water. Immerse 20mm of the handpiece head in the water then operate the
handpiece for about 15seconds. (Fig. 6)
3) Clean debris from the surface of the handpiece with clean water.
This icon denotes that the product can be washed via Thermo Disinfector.
Refer to the Thermo-Disinfector manual.
*Use a Thermo-Disinfector after the procedures 1) and 2).
CAUTION
• After washing with Thermo-Disinfector and prior to lubrication, dry the product until all internal moisture is thoroughly removed.
Thermo-Disinfector moisture remaining inside the product could reduce the effect of lubrication and could cause corrosion inside of the
product.
• To clean the product never use any solvent such as benzine or thinner.
Fig. 6
9

Publicidad

Tabla de contenido
loading

Tabla de contenido