Connecting the power
pack and charging the
battery block
4.4 Setting up and aligning the scale
58 •
WARNING!
Personal injury and damage to device due to use
of incorrect power packs
The voltage supplied by standard power packs
may be higher than their indicated rating. This
may cause the scale to overheat, catch fire,
melt or short-circuit.
►
Only use genuine seca plug-in power packs
with a controlled 12 V output voltage.
1. Insert the power plug necessary for your power
supply in the power pack.
2. Insert the device connector of the power pack in
the connecting socket in the base of the scale.
3. Plug the power supply unit into a power supply
socket.
4. Leave the scale connected to the power supply for
at least 24 hours when charging for the first time in
order to charge the battery block fully.
The scale is fully assembled on delivery.
NOTICE!
Incorrect measurement due to force shunts
If the housing of the scale is resting on some-
thing, e. g. a towel, the weight will not be meas-
ured correctly.
►
Set up the scale so that it is only in contact
with the floor with the device feet.
1. Place the scale on a flat, stable surface.
NOTICE!
Incorrect measurement due to incorrect
alignment
The spirit level is very delicate. Any additional
weights, such as towels, will result in incorrect
alignment of the scale.
►
Only align the device without a load.