OPERATING ISSUE
Error 8.1
Error 8.2
Error 8.3
Error 8.4
Err 9.0
Err 9.5
Err 53
CAL E
LOW.rEF
REF.WT Err
If the troubleshooting section does not resolve your problem, contact Uline Customer Service at 1-800-295-5510.
NOTE: Before using the indicator for the first time, the internal rechargeable battery should be fully
charged for 12 hours. The indicator can be operated during the charging process. The battery is
protected against overcharging and the indicator can remain connected to the AC power line.
π
CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA
PAGE 8 OF 24
TROUBLESHOOTING CONTINUED
CAUSES
Weight reading exceeds Power On Zero
limit.
Weight reading below Power On Zero limit.
Weight reading exceeds Overload limit.
Weight reading below Underload limit.
Internal fault.
Calibration data not present.
EEPROM data incorrect.
Calibration Error. Calibration value outside
allowable limits.
The average piece weight of the parts is
small (warning).
The average piece weight of the parts is
too small.
1-800-295-5510
uline.com
RECOMMENDATIONS
Remove load from scale. Recalibrate
scale.
Remove red shipping tags from under pan.
Add load to scale. Recalibrate scale.
Reduce load on scale.
Add load to scale. Recalibrate scale.
Service required.
Calibrate scale.
Service required.
Repeat calibration using correct
calibration weights.
Use parts with average piece weight
greater than or equal to 1 division.
Use parts with average piece weight
greater than or equal to 0.1 division.
0318 IH-2587