APPLICATION MODES

Only modes enabled in the mode menu are displayed.
WEIGHING
1.
Place item to be weighed
on the scale. The illustration
indicates a sample of
1.5 kg., gross weight.
NOTE: To return to the
weighing mode from
the parts counting
mode, press and hold the MODE button until
WEIGH is displayed.

PARTS COUNTING

1.
Use this mode to count
parts of uniform weight.
The indicator determines
the quantity based on the
average weight of a single part. All parts must be
uniform in weight for accurate measurements.
2. To enter the parts counting mode, press and hold
the MODE button until count is displayed.
AVERAGE PIECE WEIGHT (APW)
When the MODE button is
CLr.PW
released,
Pcs is
displayed.
NOTE: If no APW has
been previously stored, the CLr.PW display is
skipped and the display shows PUt 10 Pcs.
CLEARING A STORED APW
Press the YES button to clear the stored APW.
RECALLING A STORED APW
1.
Press the NO button to
recall the existing APW.
2. Press the FUNCTION MODE
button to temporarily
display the APW value.
PAGE 5 OF 24
OPERATION CONTINUED
ESTABLISHING THE AVERAGE PIECE WEIGHT (APW)
The display shows PUt 10 Pcs.
ESTABLISHING A NEW APW
1.
2. To establish the APW,
BEGIN COUNTING
1.

LEGAL FOR TRADE

When the indicator is used in
trade or a legally controlled
application, it must be set up,
verified and sealed in accordance with local weights
and measures regulations. It is the responsibility of the
purchaser to ensure that all pertinent legal requirements
are met.
SETTINGS
Before verification and sealing, perform the following steps:
1. Verify that the menu settings meet the local weights
2. Perform a calibration.
3. Set Legal for Trade to ON in the setup menu.
4. Exit the menu.
5. Disconnect power from the indicator and open the
6. Set the security switch to the ON position.
7.
8. Reconnect power and turn the indicator on.
Press the NO button to
increment the sample size.
Choices are 5, 10, 20, 50,
100 and 200.
place the specified
quantity of samples on
the scale and press the
YES button to capture
the weight.
Place the parts on the
scale and read the count.
If a container is used, be
sure to tare the empty
container first.
and measures regulations.
housing.
Close the housing.
0318 IH-2587