4.5 Sensor setup menu (continued)
Bypass
Test/Maintenance
Reset
Test/Maintenance
Sensor information
Calibration history
English
Yes/No (plus and clear)
Yes: Setting for bypass application (inhibits wiper "extension")
Counter for customer service settings: 0–1000 days (180 days are recommended)
Check the service contract and enter the defined value (number of days).
0 = Service deactivated
The instrument resets the settings to the default configuration.
Parameter: eco: NO
-N; plus, clear: NO
x
Unit: mg/l
Interval: 5 min
Response time
: eco, plus: 3 Counts; clear: 1 Count
Wiper mode: eco: Single ; plus, clear: B-A-B, B-A-B
Select NITRATAXplus/eco/clear, Location, Serial number, Range, Path length, Wiper part
number, Model number, Software version, Driver version, Production date
Name of connected sensor: NITRATAX plus/eco/clear
Location
Serial number: serial number of connected sensor
Range: Measuring range corresponding to the measuring path
Path length: Width of the measuring path
Wiper part number: Item number
Model number: Item number
Software version: Sensor software
Driver version: Structure, Firmware, Content
Production date
Overview of Offset, Factor, Date, Dext 100%, Dext 50%, Dext 25%, Factory calibration, R , M, IR
and IM
Offset: Adjustable on the Calibration menu
Factor: Adjustable on the Calibration menu
Date: Date of the last change of Offset and/or Factor
Internal calibration data:
Dext100%
Dext 50%
Dext 25%
Factory calibration: Internal calibration data
R: Internal calibration data
M: Internal calibration data
IR: Internal calibration data
IM: Internal calibration data
3
Operation
7