ATTENTION
Use thermal protection gloves.
The distillation tube is above 60ºC
once distillation is finished.
Note:
To assure that the automatic stop sys-
tem will work properly:
The front side adjustable legs must be
adjusted to left the unit slightly up from
the front side.
J
.P. SELECTA s.a.u.
ENE/2020
5 Operation
The control panel allows function selection and shows the distiller alarms status
Once the unit is plugged in, put it on, by pressing the main switch. The unit goes to
stand by status (ON).
Note: If main switch is pushed consecutively with an interval of less than 3 seconds,
the unit could start in abnormal mode. So wait for a few seconds to restart the unit.
On starting the unit for the first time, the water pump starts to work, automatically,
to fill the steam generator. This operation usually takes near 40 seconds and it will
stop when the steam generator is filled. Restart the unit if the alarm blocks it.
If, after this 40 seconds, this operation is approx not finished, please check that
the water tank (visual check on the right hand side level), and also check the water
conductivity. (See 4.4)
5.1 Insert/ Remove the distillation tube
• Insert the distillation tube starting from the PTFE tube.
• Push down the fixing lever.
• Check the upper black rubber closing and the tube fixing.
Proceed in a similar way to remove the distillation tube.
5.2 Place the collector flask.
The collector flask must be placed carefully to assure that the automatic stop system
will work properly.
The flask wall must be in contact with the level sensor face.
Placing 200ml graduated flask
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso)
Check properly
close
Check
fixing
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 26
Placing the Erlenmeyer