Chapter 6: Maintenance
6.2

Reassembling the hand sealer

1
1
16
Before reassembling the hand sealer, lubricate the following movable
parts using vaseline:
– the leaf spring near the bolt
– the sealing head inside the guiding range
– the bar in the lever
– all sliding parts of the hand sealer housing
2
Fresenius Kabi CompoSeal Universal OP-EN 6/07.13
Put down the hand sealer. The lever must be
in the position shown in the illustration.
Fully insert the leaf spring into the hand
sealer (1).
Place the sealing head onto the hand sealer
(1).
Close the lever (2).
The lever will click into place.