9 Error messages
Error 1 Pressure sensor failure or overpressure.
Error 2 PT100 probe failure or overtemp.
Error 3 External alarm. Pressure switch or safety thermostat trigered.
To reset the safety thermostat, press its button num-
ber 20.
The thermostat is adjusted in the factory, it must not be
operated except to reset it.
Error 4 Open door interlocking during the cycle.
Error 5 Open door during a cycle.
Error 6 Pressure over +0.05bar with the autoclave stopped.
Error 7 E2PROM failure.
Error 8 Out of limits temperature during sterilization.
Error 9 Mains supply failure during cycle.
10 Spare parts
07292
RS-232 communication wire
07315
26 pins communication wire
10044
R4 closing
13006
Contactor
15410
Fuse 16 A
16055
Solenoid electromagnet
16057
Electrovalve NA 3 mm 24 V
16212
Pressure switch
20096
Bipolar switch
29358
Frontal circuit
29459
Power circuit
36027
Solid relay 25A
39001
Heating element 2800W/230V
43031
Thermostat
43071
Pt100 probe
11 Lid emergency opening
In case of power supply failure, it is possible to open the lid manually.
Attention: Before opening the cover by using this system, make
sure that the gauge pressure is 0 bar.
Remove the top drawer from the display by pulling hard (it is not screwed
down, but snapped). A yellow cap will appear, gently pull it until you
hear a click. The closure will be opened.