12 Recommendation to achieve a perfect
sterilization
The material to be sterilized needs to be clean, free from any encrusted
or residual material. We recommend to wash it with a good detergent
and distilled water, then rinse it with distilled water.
Do not overload the baskets, drums or trays, always leave a space
of 1 or 2 cm between items to allow a free passage of vapour and to
ease drying.
12.1 Sterilization of instruments in bulk
Place the instruments on a tray with waterproof paper leaving a space
between each instrument.
Try to keep similar material instruments type together on a tray.
12.2 Tubing
The tubing needs to be clean and drained. Place the tubing so that
the openings on both ends are open. Make sure that the tubes are
not touching the autoclave wall, blocked or twisted and that nothing
can block the inside.
12.3 Vessels
Never put hermetically sealed vessels in the autoclave.
Place the contents open end down to prevent any residual water being
left inside.
12.4 Liquids
Place the receptacle with the liquid inside on a drip tray, making sure
that the receptacle can withstand the temperature.
Do not fill the receptacle more than 2/3rds of its maximum capacity,
DO NOT CLOSE THE LID IF IT HAS A HERMETICAL SEALED, only use
a lid if the pressure can be released.
Try to avoid using narrow neck vessels.
At the end of the cycle, let the autoclave cool down slowly (until the
pressure is 0 kg/cm²).
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.jpselecta.es
FEBRUARY 2020
Tel 34 937 700 877 Fax 34 937 702 362
Page: 27

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Presoclave iii 80l40017584001759

Tabla de contenido