Conformidad y licencias
to EC-Type Examination Certificate No. FM14ATEX0084X
13
Description of Equipment or Protective System:
The YCO14-Z Ex-Link Converter is a dedicated device providing power and data to the PMA.Evolution
EVO1X or LAB1X weighing platforms (FM14ATEX0085X). All signal outputs to the PMA.Evolution are
available in RS232 communications. The YCO14-Z Ex-Link Converter is housed within non-metallic
enclosure. A dedicated potential equalization connection is provided on both the PMA.Evolution and the
YCO14-Z Ex-Link Converter. Power and data connections between the YCO14-Z and the
PMA.Evolution are made via a multi-conductor CAT6 cable.
Electrical ratings
Um = 250 V
14
Specific Conditions of Use:
1. The YCO-14Z Ex-Link Converter shall only be used with the PMA.Evolution (FM14ATEX0085X).
2. The protective ground connectors on the PMA.Evolution and the YCO14-Z Ex-Link Converter shall be
connected together using a potential equalisation conductor with a cross-sectional area of at least 4 mm
3. The CAT6 cable used between the PMA.Evolution and the YCO14-Z Ex-Link Converter shall be obtained from
Sartorius.
15
Essential Health and Safety Requirements:
The relevant EHSRs that have not been addressed by the standards listed in this certificate have been identified
and assessed in the confidential report identified in item 8.
16
Test and Assessment Procedure and Conditions:
This EC-Type Examination Certificate is the result of testing of a sample of the product submitted, in
accordance with the provisions of the relevant specific standard(s), and assessment of supporting
documentation. It does not imply an assessment of the whole production.
Whilst this certificate may be used in support of a manufacturer's claim for CE Marking, FM Approvals Ltd
accepts no responsibility for the compliance of the equipment against all applicable Directives in all
applications.
This Certificate has been issued in accordance with FM Approvals Ltd's ATEX Certification Scheme.
17
Schedule Drawings
A list of the significant parts of the technical documentation is annexed to this certificate and a copy has
been kept by the Notified Body.
18
Certificate History
Details of the supplements to this certificate are described below:
Date
30
April 2015
th
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals Ltd. 1 Windsor Dials, Windsor, Berkshire, UK. SL4 1RS
T: +44 (0) 1753 750 000 F: +44 (0) 1753 868 700 E-mail:
F ATEX 020 (Apr/14)
54
Instrucciones de manejo EVO1X | LAB1X
SCHEDULE
Description
Original Issue.
atex@fmapprovals.com www.fmapprovals.com
.
2
Page 2 of 2