See Figure 7d for the fixing centre distances.
Wiring connection/polarity
The bracket is supplied complete with the ambulance battery connection cord (5) (already fitted), which
powers the device and charges its battery when it is connected to the bracket.
Connect the wires to the ambulance following the polarity shown in Figure 7e.
If the cord is disconnected from the bracket for any reason, reconnect it following the polarity shown in
Figure 7f.
N.B.: Take great care over the polarity (+ and –) because if it is reversed the device's electronic circuit
board and battery will be damaged beyond repair.
PROBLEMS
Excessive noise.
The unit switches on but does not aspirate. - Damaged pump.
The vacuum rate cannot be regulated. Damage to the internal hydraulic system
When the appliance is switched on, the
protection fuse always trips.
The vacuum gauge does not work. Liquid penetrating the pneumatic circuit. Send to the assistance circuit.
Note: if you experience faults or malfunctioning problems different to those listed above, always and
exclusively contact authorised assistance centres.
Professional portable medical device for secretion removal, suitable for fixed and transportable use.
Complete with flame retardant ABS V0 casing and piston-cylinder electrical compressor lubricated for life.
Risk class under Directive 93/42/EEC:
Suction class:
Protection level against entry of liquids and solids:
Voltage:
Absorption:
Adjustable vacuum level:
Vacuum gauge precision class:
Air flow:
Temporary use:
Dimensions:
Weight:
Noise level:
Multi-voltage switching battery charger:
Internal battery:
Battery charge duration:
Operating conditions:
Temperature:
Air humidity:
Storage conditions:
Temperature:
Air humidity:
Operating-storage conditions:

PROBLEMS, CAUSES AND SOLUTIONS

POSSIBLE CAUSES
Damaged pump or blockages in the
internal aspiration circuit.
- Vacuum regulator fully open. Connection
tubes disconnected and/or badly
connected, broken connection tubes.
Container not in a vertical position, full, or
defective overflow valve. Possible
blockage of the hydraulic circuit inside the
unit.
or blockage of the connection tubes to
the aspiration unit.
Pump probably damaged or in shortcircuit. Send to the assistance circuit.

TECHNICAL CHARACTERISTICS

Send to the assistance circuit.
- Send to the assistance circuit.
- Check the position of the vacuum
regulator. Check the connections and the
integrity of the tubes. Position the
container in a vertical position, check the
overflow valve (blocked) and/or replace
the silicon tubes.
Send to the assistance circuit.
IIb
High vacuum / High flow
IP44D
12VDC
3A
from 0 to –0.85 bar
2.5% under UNI EN 837
30 lt/min operating without restrictions +/– 10 %
max. 60 minutes
38 (length) x 13.5 (width) x 26 (Height) cm
4.7 kg approx
61dBA
PRI: 100 ÷ 240V~ 50 – 60Hz
SEC: 14VDC 4.28A
12VDC 4Ah Lead, hermetic
45 minutes at maximum suction power
min. 0° C; max 40° C
min. 10 %; max 95 %
min. –10° C; max 50° C
min. 10 %; max 95 %
min. 690 hPa; max 1060 hPa