General Technical Specifications - Dräger FG7000 Guía Rápida De Instrucciones

Ocultar thumbs Ver también para FG7000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

General technical specifications

Display:
Interfaces:
Power supply:
Battery life:
Dimensions:
Weight:
Operating temperature:
Storage temperature:
Relative humidity:
Air pressure:
Certification:
Display
Combustion air
temperature
Exhaust gas tem-
perature
O
2
CO
NO
(1)
Draft
(2)
Pressure
(3)
Medium pressure
(4)
(ext. sensor, option)
High pressure
(5)
(ext. sensor, option)
*MV = Measuring value **MR = Measuring range
(1) NO sensor only available in NO configuration
(2) = Pmax. 1,250 Pa
(3) = Pmax 750 hPa (hPa (mbar))
CO, undiluted
CO
, carbon dioxide
2
Exhaust gas losses
Efficiency
CO/CO
ratio
2
* = taking into account of the gain in calorific value
Colour display with touch screen
USB, IR
Li-ion battery, 3.6 V, 3,400 mAh, charge level indicator,
Primary charger 100 - 240 V AC; Secondary 5 V DC, 1 A
Typical 8 hours (depending on type of measurement and the adjusted dis-
play backlight)
93 x 225 x 40 mm (W x H x D)
approx. 540 g
+ 5 °C ... + 40 °C
- 20 °C ... + 50 °C
10 - 90 % RH, not condensing
800 to 1,100 hPa
DIN EN 50379 Part 1 and Part 2
Measuring range
- 10 ... + 100°C
0 ... + 600 °C
0 ... 25 Vol %
0 ... 8,000 ppm
0 ... 2,000 ppm
- 500 ... + 500 Pa
0 ... 100 hPa (hPa
(mbar))
+ 101 ... 160 hPa (hPa
(mbar))
-100 ... +3,500 hPa
(hPa (mbar))
0.01 MPa ... 2.5 MPa
(0.1 ... 25.00 bar)
calculated
0 ... 9,999 ppm
calculated
0 ... CO
0 ... + 100 %
calculated
- 20 ... + 100 %*
0 ... + 100 %
calculated
0 ... + 120 %*
calculated
1.00 ... 9.99
Resolution
Accuracy
0.1 °C
± 1 °C
0.1 °C (< 100 °C)
± 2 °C or 1.5 % of MV*
1 °C (≥ 100 °C)
0.1 Vol %
± 0.3 Vol %
1 ppm
0 ... 200 ppm:
± 10 ppm or 10 % of MV*
201 ... 2,000 ppm:
± 20 ppm or 5 % of MV*
2,001 ... 8,000 ppm:
± 100 ppm or 10 % of MV*
1 ppm
0 ...600 ppm:
< ± 5 ppm or 5 % of MV*
0.1 Pa
- 50 ... + 200 Pa:
± 2 Pa or 5 % of MV*
0.01 hPa (hPa
0.5 hPa (hPa (mbar)) or 1 % of
(mbar))
MV*
0.1 hPa (hPa
5 % of MV*
(mbar))
1 hPa (hPa (mbar))
< 1 % of MR**
0.001 MPa (0,01
< 1 % of MR**
bar)
(4) = Pmax 4,000 hPa (hPa (mbar))
(5) = Pmax = 3.5 MPa (35 bar)
1 ppm
max.
0.1 Vol %
2
0.1 %
0.1 %
0.01
6

Publicidad

Tabla de contenido
loading