INSTRUCTION MANUAL CODE 80256 REV 0
42
5
43
24
J
.P. SELECTA s.a.
January/2011
ASSEMBLY REFRIGERANT GLASS
•
Place the P.T.F.E. support of the gaskets making the anti-rotating pin
meet the hole shown on the picture.
•
Part placed with the gasked.
•
Fix the cooling glass with the stopper.
ASSEMBLY COLLECTING CONDENSED FLASK
•
To assemble the sample flask lock the metallic hock as shown on the
picture and tighten the thread until the glass becomes fixed.
INSTALLATION OF THE PROBE OF THE COOLING GLASS
•
Insert the probe (5) into the lateral hole of the refrigerant glass keeping
in mind that it has to fit well in order to avoid vacuum leakages.
•
Connect the probe to the plug (24).
WATER COOLING GLASS CONNEXION
•
Connect the tap water with a silicone pipe to the inlet of the cooling
glass (42).
•
Connect a silicone pipe to the outlet of the cooling glass (43) and then
give it to the waste.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
Tel 34 937 700 877 Fax 34 937 702 362