PULOX PO-250 Manual De Instrucciones página 87

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Appendix 1
Guidance and manufacturer's declaration – electromagnetic
emissions- for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emission
The CMS50D+ is intended for use in the electromagnetic environment
specified below. The customer of the user of the CMS50D+ should
assure that it is used in such and environment.
Emission
test
RF
emissions
CISPR 11
RF
emission
CISPR 11
Guidance and manufacturer's declaration – electromagnetic
immunity – for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The CMS50D+ is intended for use in the electromagnetic environment
specified below. The customer or the user of CMS50D+ should assure
that it is used in such an environment.
Immunity
test
Electrostatic
discharge
(ESD)
IEC 61000-4-
2
Power
frequency
(50/60Hz)
Compliance
Group 1
Class B
IEC
60601
test
level
6 kV
contact
8 kV
air
3A/m
Electromagnetic environment –
guidance
The CMS50D+ uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The CMS50D+ is suitable for use in all
establishments other than domestic
and those directly connected to a low
voltage power supply network which
supplies buildings used for domestic
purposes.
Compliance
level
6 kV
contact
8 kV air
3A/m
Electromagnetic
environment -
guidance
Floors should be
wood, concrete or
ceramic tile. If floor
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Mains power quality
should be that of a
typical commercial

Publicidad

Tabla de contenido
loading

Tabla de contenido