magnetic
field
IEC 61000-4-
8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration – electromagnetic
immunity – for EQUIPMENT and SYSTEMS that are not LIFE-
Guidance and manufacturer's declaration – electromagnetic immunity
The CMS50D+ is intended for use in the electromagnetic environment
specified below. The customer or the user of CMS50D+ should assure
that it is used in such an environment.
Immunity
test
Radiated
RF
IEC
61000-4-
3
SUPPORTING
IEC
60601
Compliance
test
level
level
3 V/m
80
MHz
3 V/m
to 2.5
GHz
or hospital
environment.
Electromagnetic
environment - guidance
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the CMS50D+,
including cables, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
3
5 .
d
P
V
1
3
5 .
d
P
E
1
80 MHz to 800 MHz
7
d
P
E
1
800 MHz to 2.5 GHz
Where P is the maximum