for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the CMS50D+
The CMS50D+ is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the
user of the CMS50D+ can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the CMS50D+ as
recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above,
the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Appendix 2
State
Low voltage alarm
SpO2 alarm
Pulse rate alarm
Probe error alarm
Separation distance according to
frequency of transmitter (m)
80 MHz to 800 MHz
d
0.12
0.37
1.17
3.69
11.67
Alarm condition
delay
1s
330ms
330ms
16ms
3
5 .
P
E
1
Alarm signal
generation delay
20ms
20ms
20ms
20ms
800 MHz to
2.5 GHz
7
d
P
E
1
0.23
0.74
2.33
7.38
23.33