Enraf Nonius Radarmed 950+ Instrucciones De Utilizacion página 54

Ocultar thumbs Ver también para Radarmed 950+:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Select by means of the "Central control" [8] the desired
indication.
Operate the "Accept key" [7], the unit settings and advise
information appears on the display. The program number is
displayed in the upper right corner [12].
Set the power.
Treatment starts as soon as the power is adjusted.
To retrieve by number
Operate "Program key protocols" [3]. The display shows the
first cluster.
Operate "Program key protocols" [3] once again. In the
upper right corner appears a flashing letter with besides it a
number (e.g. A1). The parameter settings concerning that
program aredisplayed on the display.
Select by means of the "Central control" [8] the desired
cluster (A till E).
Move the cursor one position to the right by means of the
"Display parameter selectors" [4]. Now the number will
flash.
Select by means of the "Central control" [8] the desired
program number. The parameter settings and advise
information are displayed.
Operate the "Accept key" [7].
Set the power.
Treatment starts as soon as the power is adjusted.
EN109-1435752-42
ENGLISH
Page 22 of 29

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido