Cleaning, Disinfection and Sterilization
Operation
Manual
Cleaning or
4
assisted by an
ultrasonic
device
END-IFU-PROPEX-PIXI-V15- WEB-EUR-Multilingual -2020-11
Operating mode
•
Place the pre-cleaned
accessories into the
cleaning bath for the
prescribed contact
time (e.g.
®
CIDEZYME
,
®
ENZOL
Enzymatic
Detergent Solutions,
Johnson & Johnson
Medical, 0.8% for 1
minute); make sure
that the products are
fully immersed (if
necessary, use a soft
brush to carefully
brush them down).
•
Remove the
accessories from the
cleaning bath and
rinse them off
thoroughly at least
three times for a
minute each time with
sterile, deionized
water.
•
Next, place the
accessories in an
ultrasonic bath with a
cleaning agent (e.g.
®
CIDEZYME
,
®
ENZOL
Enzymatic
Detergent Solutions,
Johnson & Johnson
Medical, 0.8% for 20
minutes).
Warning
•
Make sure that the
accessories are not in direct
contact with one another.
•
The hook mechanism has to
be activated during the
cleaning process, the rinsing
process and the ultrasonic
bath cleaning (press several
times the push button) to
allow the inner parts to be
cleaned more effectively.
•
Make sure that the products
are fully immersed (if
necessary, use a soft brush
to carefully brush them
down).
•
No visible impurities should
be observed on the
accessories.
27/222