Chopin Aqua-TR Modo De Empleo página 85

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

User's manual 
07/2015
Aqua-TR
■ Calibrations
❏ This menu allows to modify some of the calibrations
(unapproved) and to enter new calibrations.
❏ Click on "Meas. and calibrations" and choose "Calibrations".
❏ Only unapproved calibrations ( "approved" box unticked) can be
modified.
A calibration is form:
H2O
=
A0
+
A1 Lin
Lin : rough measurement obtained with the linear calibration
❏ To modify a calibration, double-click on the box to be modified,
enter the modification and click on "PC => Aqua-TR" to update the
Aqua-TR memory.
❏ To insert a new calibration :
- Click on "Insertion", enter the name of the new calibration and click
on "Validate". The name of the new calibration must not have more
than 2 consecutive characters in common with an approved
calibration.
- Enter the parameters of the new calibration (A0, A1, A2 and A3),
the range of the calibration (Mmin, Mmax), the HLW ref. and the
HLW. Entering Mmin,Mmax, HLW ref. and HLW is optional.
- Click on "PC => Aqua-TR" to transfer the new calibration.
Note:
For any modification on approved calibration, contact your Chopin
Technologies entitled repair organization.
2
Lin
---------- -
+
A2
+
A3
100
GB
3
Lin
--------------- -
10000
85