GB
Drawer problem.
Err 15.
❏ Check if drawer is well inserted. If problem remains, contact your agent.
Trap problem
Err 16.
❏ Check if grains or any visible object prevents the inlet trap to close ; start again test
with the vibration option selected (see
Level detection problem.
❏ Check the hopper for dirt.
Err 17.
❏ Deactivate the option if not used.
❏ Otherwise contact your agent.
Measuring not possible, excessive grain moisture.
Err 18.
❏ Run a test with different grain. If problem remains, contact your agent.
Ambient temperature out of range.
Err 19.
❏ If ambient temperature is however in the range 2 -45 °C, contact your agent.
Gain temperature out of range.
Err 20.
❏ If grain temperature is however in the range 2 -45 °C, contact your agent.
Temperature difference out of range (> 20°C).
❏ Start again test with grain at ambient temperature. If problem remains, contact your
Err 21.
agent.
Moisture result out of range.
Err 22.
❏ Start again test with different grain. If problem remains, contact your agent.
Insufficient grain level.
Err 23.
❏ Add grain in the hopper and press again on "TEST".
No response from measure microcontroller.
Err 24.
❏ Contact your agent.
Motor 1 problem (hopper trap).
Err 25.
❏ Contact your agent.
Motor 2 problem (cell traps).
Err 26.
❏ Contact your agent.
TR Part No.
CHOPIN part No.
06200537
NLC 28
06202939
TM3 66
06201092
TM3 42
06200566
TM3 37
92
Aqua-TR
page
67) ; if problem remains, contact your agent.
4.3 Recommended replacement frequency of parts
that have an influence on the result:
Description
Periodicity
Cell traps motor
Hopper motor
Temperature probe
Display seal
(*) Replacement of some of these parts needs to open the
equipment. This shall be carried out by a CHOPIN technologies
entitled technician. Please contact you agent.
User
2 years
2 years
2 years
2 years
Agent (*)
X
X
X
X
User's manual