12. Specifications
Process connection
Type of process
connection
EN
Protection tube
Protection tube
diameter
Insertion length U
Material (wetted)
1) Not for protection tube diameter 3 mm [0.12 in]
If the resistance thermometer is to be operated in an additional protection tube, a
spring-loaded compression fitting must be used.
Output signal (4 ... 20 mA version)
Analogue output
Load R
A
28
G ¼ B
G ⅜ B
G ½ B
¼ NPT
½ NPT
M12 x 1.5
M20 x 1.5
3 mm [0.12 in]
6 mm [0.24 in]
8 mm [0.31 in]
50 mm [1.97 in]
1
75 mm [2.95 in]
100 mm [3.94 in]
120 mm [4.72 in]
150 mm [5.91 in]
200 mm [7.87 in]
250 mm [9.84 in]
300 mm [11.81 in]
350 mm [13.78 in]
400 mm [15.75 in]
Other insertion lengths on request
Stainless steel 1.4571
4 ... 20 mA, 2-wire
R
≤ (U
- 10 V) / 23 mA with R
A
B
The permissible load depends on the loop supply
voltage.
WIKA operating instructions model TR36