Binning Commands
BINCLEAR
BINNOM
<nr1>,<nrf>
BINNOM?
<nr1>
LIMHI
<nr1>,<nrf>
LIMHI? <nr1> Returns the upper limit of Bin <nr1>.
LIMLO
<nr1>,<nrf>
LIMLO?
<nr1>
SORTON
SORTOFF
Page 34
Clears the nominal values and limits of all the bins; this
has the effect of also turning off Sort, should it be
selected.
Sets the nominal of Bin <nr1> to value <nrf>; <nr1>
can be 0 to 8 (9 is the general fail bin). Note that Bin 8
is always the minor term bin (Q, D or R).
The nominal value <nrf> relates to the function
selected at the time the first bin is defined; further bins
defined relate to the same function. Selecting Sort will
force that selected function.
If no nominal value is set for a bin, the nominal value
for the next lowest bin will automatically be used. The
lowest numbered active bin must have its nominal
value set; Bin 0 must always be set for binning to be
enabled.
Returns the nominal value of bin <nr1> in the form
<nrf><rmt>.
Sets the upper limit of Bin <nr1> to <nrf>%. The upper
limit must be set before the lower limit.
Sets the lower limit of Bin <nr1> to <nrf>%. The lower
limit must be set below the upper limit (which must
have been set first). If no lower limit is set the
instrument will use the negative of the upper limit, i.e.
the limits will be symmetrical about the nominal.
Returns the lower limit of Bin <nr1>.
Note: Limits may be set for bins with no nominal
value; the nominal used will be that of the next
lowest bin which has a nominal set.
Enables binning (sort). Enabling sort forces the
measurement function associated with the binning set-
up. Sort can only be enabled if at least one bin has
been defined.
Disables binning (sort).