3.11 VlowZ
The Vlow Z position measures AC voltage with lower impedance (500kΩ) than
normal voltage measurement, this function is used to avoid measuring ghost voltages.
4. MAINTENANCE
The instrument has no parts that can be replaced by personnel who are not trained
and approved. Any non-approved repair or other work, or replacement of a part by
an "equivalent", may severely compromise safety
4.1 Cleaning
Periodically wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents. Dirt or moisture in the terminals can affect the readings. Dry the
meter perfectly after cleaning and before use.
32

4.2 Testing the 10A Fuse

To avoid electrical shock, remove the test leads and any input
signals before replacing the fuses.
1. Turn the rotary switch to
position and toggle the
yellow key.
2. Plug a test lead into the + terminal and touch the probe to
the A or mA/μA according to the fuse to be tested
A good A terminal fuse is indicated 000.0Ω to 000.2Ω.
3. If the display reads OL, replace the fuse and test again
4. If the display shows any other value, have the meter
serviced.

4.3 Replacing the batteries and fuses

To avoid false readings, which could lead to possible electric
shock, or personal injury, replace the batteries with LR6
batteries as soon as the battery indicator
appears.
To prevent damage of injury, install ONLY replacement fuses with the specified
amperage, voltage, and interrupt ratings.
Disconnect test leads before opening the battery door.
F1: Fast Fuse 630mA 1000V 10kA size: 6.3 x 32, UK standard
F2: F2: Fuse 10A 1000V 30kA size: 10 x 38, UK standard

4.4 Metrological check

Like all measuring or testing devices, the instrument must be checked regularly. This
instrument should be checked at least once a year. For all checks and calibrations,
contact a Würth branch or give the instrument to your Würth sales representative.
4.5 Service, and Parts
For all repairs before or after expiry of warranty, please return the device to a Würth
branch or give it to your Würth sales representative without having dismantled it.
5. WARRANTY
Except as otherwise stated, our warranty is valid for 2 years starting from the date
on which the equipment was sold. Extract from our General Conditions of Sale
provided on request. The warranty does not apply in the following cases:
• Inappropriate use of the equipment or use with incompatible equipment
• Modifications made to the equipment without the explicit permission of the
manufacturer's technical staff
• Work done on the device by a person not approved by the manufacturer
• Adaptation to a particular application not anticipated in the definition of the
equipment or not indicated in the user's manual
• Damage caused by shocks, falls, or floods.