Locking screw
2
Binocular tube
3
Eyepieces
Fig. 18
Binocular tube and
eyepieces
Operation
Caution
Only connect cables and sockets that are the same colour and size.
Never force connections.
• Install the remote interlock contact. If you do not wish to install an
interlock, connect the yellow coding plug provided in place of the
interlock (yellow connector, see VISULAS 532s user manual).
• The power cable for the IT 4L should be connected to a fixed power
socket.
• Connect the VISULAS YAG III to one of the sockets on the
instrument table (see separate IT 4L user manual). The VISULAS 532s
must be plugged into a separate fixed earthed socket.
• Remove the protective cap from the optical fibre of the applicator.
Caution
Never touch the end of the optical fibre with your fingers. Protect the
end of the optical fibre from contamination.
Because of the laser's high power output, any dirt may lead to damage
to the end surface of the fibre.
• Insert the optical fibre carefully into the socket for the optical fibre
on the VISULAS 532s console (see VISULAS 532s user manual) and
screw in firmly.
Caution
Screw in by hand only. Do not use tools!
• Fold out the control panel foot, unwind as much of the control
panel cable as required and connect it to the front of the instrument
(see also VISULAS 532s user manual).
• Adjust the foot to the required angle and place the control panel in
the location required.
Note:
An additional component (e.g. an optical splitter) can be
mounted between the binocular tube and the microscope
body. This assembly is locked by the locking screw (1).