4. Assembly
4.1 Wall mounting
Wall mounting BASE-STATION 3 NG for LIFE-BASE 4 NG WM 8195
Warning!
•
Only the LIFE-BASE III and LIFE-BASE 4 NG portable systems may be sus-
pended in the BASE-STATION 3 NG wall mounting. The weight and di-
mensions of these portable systems are exactly matched to the wall
mounting, so use of other portable systems is not permitted.
•
Hanging up third-party products or adapters from other manufacturers
is not permitted, as this does not comply with the use in accordance with
the purpose of this wall mounting.
Fitting steps
WM 8195
Note
•
The surface for attachment must be sufficiently strong to hold the wall
mounting and the portable system securely to EN 1789 (ambulances)
and/or RTCA DO 160E.
•
The charger interface connection to the 12 V on-board electricity supply
must be fused with an 8 A fuse.
1. Look for a suitable place to attach the wall mounting,
somewhere plane and vertical.
2. Mark all 9 bores at the mounting point.
3. Drill the holes using a 4.2 mm drill bit.
4. Insert nine M5 threads in the surface for attachment.
5. Screw on the wall mounting using the 9 screws en-
closed. The portable system must be fitted horizontally.
6. Connect the charger interface to the 12 V on-board
electricity supply of the emergency vehicle using the
plug connector for vehicles.
EN
Assembly