Hanging From A Stretcher Or A Hospital Bed - Weinmann LIFE-BASE 4 NG Descripción Del Aparato E Instrucciones De Manejo

Sistema de transporte
Ocultar thumbs Ver también para LIFE-BASE 4 NG:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Remove portable system

5.2 Hanging from a stretcher or a hospital bed

You can also suspend the portable system from a stretcher, a hospital bed or a standard
rail, as long as the tubing diameter does not exceed 35 mm.
Warning!
During the journey/flight, the portable system must be suspended in the wall moun-
ting provided for it, as otherwise the portable system may fall over or be thrown
around the vehicle.
Caution!
Note that the portable system may not be loaded with any more than twice the
weight of the portable system when suspended (e.g. from a hospital bed).
1. For unlatching, it is essential to grasp through the han-
dle so that the portable system does not fall out of the
wall mounting and injure someone.
2. To unlatch, pull the catch cross-bar upwards.
3. Swivel the portable system forwards a little way and lift
the portable system up off the bottom rail.
1. Undo the knurled screws.
2. Extend the hanger the required distance.
3. Suspend the portable system from a stretcher, a hospi-
tal bed or a standard rail and tighten the knurled screws
back up.
Note
LIFE-BASE 4 NG can only be hung back in a wall
mounting once you have pushed the hanger back
into its original position.
EN