Descargar Imprimir esta página

Invacare Aquatec Ocean E-VIP Manual Del Usuario página 31

Ocultar thumbs Ver también para Aquatec Ocean E-VIP:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 149
Malfunction
Emergency lowering mechanism is not
working
Tilted position of seat frame
The seat frame does not move upwards
or
The seat frame does not move
downwards
or
The seat frame does not tilt
1556395-D
Possible Cause(s)
Battery flat
Cable not fully inserted
Connector strip not connected
Faulty button
The motors were raised to different
heights
Weight limit exceeded
Cable not correctly inserted
Connector strip not connected
Cable incorrectly attached to controller
Connector cable broken
Connector cable not completely inserted
Controller faulty
Battery flat
Hand control cable is not inserted
correctly
Remedy
Charge battery, refer to 4.12 Charging
Battery, page 20, or use a replacement
battery
Insert cable completely
Connector strip connected (engaged)
Implement reset (see below table)
Hold the down button B until the drives
are lowered completely and until you
hear the limit switch click into place
Max. load 150 kg
Insert cable correctly
Connector strip connected (engaged)
Motor for tilting always on channel 7
Replace cable
Insert cable completely and close safety
bar
Implement reset (see below table)
Charge battery, refer to 4.12 Charging
Battery, page 20, or use a replacement
battery
Insert cable completely
Troubleshooting
31

Publicidad

Solución de problemas

loading