H-863, H-1231
H-4338
PRICE TAG GUN
IMPORTANT! Read all instructions and warnings
before using these tools.
TO INSTALL NEEDLE
1.
Slide needle guard over needle. (See Figure 1)
2. Put needle lock lever in unlocked position (lever
pointed to front of tool). (See Figure 2)
3. Place tool so lock lever faces away from you.
4. Insert needle by aligning side slot on needle with
guide slot on tool.
5. Turn needle lock lever to locked position (lever
pointed to back of tool). (See Figure 2)
6. Remove needle guard.
TO REMOVE NEEDLE
1.
Slide needle guard over needle.
2. Put needle lock lever in unlocked position (lever
pointed to front of tool). (See Figure 2)
3. Slide the needle out of the tool.
4. Dispose of needle in a Biohazard Sharps container.
NOTE: For contaminated needles, see safety
instructions on page 2.
TO LOAD FASTENERS
1.
Insert the end of the round connecting bar on the
fastener clip into the round guide slot at the top of
the tool. (See Figure 1)
2. Gently push the fastener clip into the guide as far as
it will go. Tool is ready for use.
PAGE 1 OF 6
1-800-295-5510
uline.com
INSTALLING AND REMOVING NEEDLES
LOADING AND ATTACHING FASTENERS
Figure 1
Figure 2
Locked Position
TO REMOVE A PARTIALLY USED CLIP
1.
Fully depress the trigger and remove the clip by
pulling upwards.
NOTE: Always depress the trigger when
removing a clip of fasteners from the tool.
Failure to do so may cause the tool to jam.
Para Español, vea páginas 3-4.
Pour le français, consulter les pages 5-6.
Guide
Slot
Needle
Lock Lever
Needle
Trigger
Needle
Guard
Unlocked Position
0421 IH-863