Siemens SITRANS TH100 Instrucciones De Servicio Resumidas página 8

Transmisores de temperatura
Ocultar thumbs Ver también para SITRANS TH100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 96
Securing the transmitter:
in the connection head cover
Transmitter
Connection head
Note
Using the lock washers (TH200/TH300 only)
The lock washers
included in the delivery are only required for securely fastening the transmitter when the transmitter
is directly installed on the round plate
Fixing screw M4x35
Round plate
3.1.2
SITRANS TH installation on DIN rail and G rail
Note
Fixing rings
The fixing rings included in the scope of delivery for the transmitter are not required for the installation on DIN rails or G rails.
You can either install the transmitter on a 35 mm DIN rail or on a 32 mm G rail. DIN EN 60715 applies to DIN rails and G
rails in this context. The DIN rail adapter required for installation can be ordered as an accessory under the Order No.
7NG3092-8KA.
Adhere to the ambient conditions specified in the technical data.
8
for a temperature probe.
in the connection head base
Ceramic base of the measuring element
Transmitter
Lock washer DIN 6799 - 3.2 A2
SITRANS TH/TR/TF
A5E32363963-01, 06/2013