Chaleco y cubre-pierna para hiper/hipotermia

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 10
Flow Check Procedure
Figure 2
When using two gray hose sets (see Figure 2), one connected to a vest and one connected to the leg wrap(s), check
for blocked water flow as follows:
1.
Disconnect one gray hose set from the machine connection. See Figure 2. If CHECK WATER FLOW light illuminates ,
check for kinks, folds or closed clamps in the other body wraps connected to the machine.
2.
Reconnect hose set to machine.
3.
Disconnect second hose from the machine. If CHECK WATER FLOW light illuminates, check for kinks, folds or closed
clamps in the other body wraps connected to the machine.
4.
Reconnect hose set to the machine.
When using only one gray hose set, connected to only the vest or leg wrap(s), it is not necessary to follow this
NOTE
procedure. The CHECK WATER FLOW light will illuminate if the flow is interrupted.
www.stryker.com
[ENGLISH]
P/N 101124 Rev D
1-3