Service And Maintenance - parkell DIGITEST 3 Instrucciones De Uso

Ocultar thumbs Ver también para DIGITEST 3:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 25
Fig. 1
E
F
A. Stimulus Adjustment / Control Button
B. Digital Display
C. Ground Clip
D. Lead Wire
KIT INCLUDES (FIG. 1):
• (1) Digitest 3 Pulp Vitality Tester
• (4) Autoclavable Stainless Steel Tooth Probes
• (1) Lead Wire
• (1) Ground Clip
• (1) High-Output Alkaline 9-Volt Battery
CLEANING AND INFECTION
CONTROL OF THE DIGITEST 3
• Autoclaving and disinfecting do not remove accu-
mulated debris. Before autoclaving or disinfecting
accessories: Rinse the accessories under warm
running water for 30 seconds to remove any exter-
nal or internal soil or debris. Using a soft soapy
cleaning brush to assist in the cleaning, if neces-
sary. Use non-ammoniated detergent or dish-
washing soap. Do not use ammoniated cleansers
or disinfectants. Rinse the item again under warm
running water for 30 seconds to remove any resid-
ual soap and blot dry with a dry lint-free towel.
Wipe device with a lint-free towel that has been
saturated in an EPA-approved, hospital-grade
intermediate or high-level disinfectant (minimum
2.5% Glutaraldehyde solution), following the
instructions for use provided by the disinfectant
manufacturer. Dry with a dry lint-free towel.
• DO NOT AUTOCLAVE THE DIGITEST 3
POWER UNIT, AS THIS WILL CAUSE
DAMAGE TO IT.
• The Power Unit should be protected using proper-
ly-sized disposable plastic barrier sleeves.
• Since the Digitest 3 incorporates sophisticated
electronic circuitry, it should not be directly
sprayed with or soaked in disinfectant. The device
may be disinfected by wiping it with a damp cloth
or paper towel moistened with an EPA-approved
high-level surface disinfectant, then wiped with
a water-damp paper towel to remove chemical
residue, and then dried with another paper towel.
• Follow the surface disinfecting protocol specifi ed
by the disinfectant manufacturer.
A
B
G
H
E. Posterior Autoclavable Probe
F. Anterior Autoclavable Probe
G. Precision Labial Autoclavable Probe
H. Precision Lingual Autoclavable Probe
C
• For further information, consult the Centers for
Disease Control website at CDC.gov, referenc-
ing the most recent version of the "Guidelines for
Infection Control in Dental Health-Care Settings".
• The Tooth Probes, Ground Clip and Lead Wire
may be sterilized in any conventional steam auto-
clave (130º-134º C for 15 min for Gravity Displace-
ment units, or 4 minutes for Pre-vacuum Cycle
units, followed by a 15-minute minimum cool
down interval). Comply with the autoclave manu-
facturer's recommended procedure, if different.
These items can be autoclaved for 250 cycles
before requiring replacement.

SERVICE AND MAINTENANCE

• Do not open the Digitest 3 Power Unit, except to
change the battery. There are no user-service-
able parts inside. Internal repairs are to be made
only by authorized Parkell personnel, by return-
ing the unit to the service address at the end of
these instructions.
• Avoid dropping the Power Unit or subjecting it to
physical shock.
• Battery should be removed if unit is to be stored
unused for more than 30 days.
• To prolong battery life, the device automatically
turns off after 14 seconds of inactivity.
• When the battery is low, the letters "LO" will blink
on the screen when the device is initially turned
on. If the battery is low, replace it as explained
in the section "Installing/Replacing the Battery".
• If a Tooth Probe becomes loose in the mount of
the Power Unit, it may be tightened by carefully
inserting a fl at screwdriver blade into the split
metal base of the probe and gently expanding
the sides apart with a twist of the tool (Figure 2).
Fig. 2
2
D