ResMed Air10 Manual Del Usuario página 6

Ocultar thumbs Ver también para Air10:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 36
Troubleshooting
If you have any problems, refer to the following troubleshooting topics. If you are
not able to fix the problem, contact your care provider or refer to
www.resmed.com. Do not try to open the oximeter adapter.
Problem/possible cause
AirSense 10 / AirCurve 10 / Lumis: Check your finger sensor or press OK to continue
without oximetry
S9: Check the oxi sensor or press dial to accept no oximetry data logging
The finger pulse sensor is not attached
to your finger
The sensor cable is not connected to the
Xpod oximeter at the start of therapy.
AirSense 10 / AirCurve 10 / Lumis: Check oximeter connection or press OK to continue
without oximetry
S9: Connect the oxi sensor or press dial to accept no oximetry data logging
The oximeter adapter is not connected
properly or data is not transmitting
correctly.
S9: Insert your SD Card or press dial to accept no oximetry data logging
The device cannot detect an SD card and
no oximetry data can be logged or stored.
4
Solution
Make sure the finger pulse sensor is properly attached
according to the instructions supplied with that
accessory.
Disconnect and reconnect the Xpod oximeter connector
and sensor cable according to the instructions supplied
with these accessories.
Remove your oximeter adapter from the device
according to the instructions. Reattach your oximeter
adapter and plug your device back into the power outlet.
Make sure that an SD card is correctly inserted in the
SD card slot of your device.

Publicidad

Tabla de contenido
loading

Tabla de contenido