Meter And Test Strip Problems - Bayer Glucometer Elite XL Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

II
See the possible solutions listed and if you need
further assistance, contact Bayer Customer Service
at: 1-800-348-8100 (24 hours a day)
MESSAGE
E-1
E-3
display
E-4
E-2
display
Display goes blank
after Test, Code or
Check Strip is inserted
into the Meter.
The Code Number
(F#) displayed is
different from the
number printed on
the Test Strip foil
packet or carton.
Check Strip test
result out-of-range.
20

Meter and Test Strip Problems

CAUSE
Meter electronics
failure.
Meter below or above
operating tempera-
ture, 50°–104° F
(10°– 40° C).
1. Strip is not fully
inserted into the
Meter.
2. Batteries are not
installed correctly.
3. Batteries are dead.
4. Too much time
has passed since
inserting the strip
and the Meter
has turned off.
1. Wrong Code
Strip used.
2. Code Strip
not inserted
properly.
1. Check Strip
(Meter end) dirty
or damaged.
2. Meter electronics
failure.
SOLUTION
Immediately contact the Customer
Service Dept. 1-800-348-8100. The
Meter will no longer function.
Allow Meter to come to room
temperature before using. If E-2
continues to display, contact
Customer Service Dept. (see above).
1. Insert strip until tab is
inside Meter.
2. Install batteries correctly.
3. Replace with new batteries.
4. Remove and reinsert the
strip.
1. Use correct Code Strip.
Discard incorrect one.
2. Reinsert Code Strip care-
fully, but firmly.
1. Wipe Check Strip with
clean dry tissue, and retest.
2. Contact Customer Service
Department.
Glucometer Elite
®
XL

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido