Continuity Test; Diode Testing - PANCONTROL PAN 118 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 67
Once the reading stabilizes, read the value. If the polarity is twisted a "Minus 
sign" is displayed. 
The  test  leads  have  an  inherent  resistance  of  approx.  0.1  ‐  0.2Ω,  which 
influences the measurement. To get a more accurate result  in the range up to 
200  Ω,  the  test  leads  should  be  shorted  before  the  actual  measurement  and 
their  resistance  value  should  be  subtracted  from  the  value  of  the  actual 
measurement. Above 1MΩ the display may fluctuate for a few seconds before 
the exact value is displayed. 
 
Continuity test 
Attention:  Before  making  any  measurements,  make  sure  the  circuit  is 
disconnected  from  any  power  source  and  all  capacitors  are  properly 
discharged! 
1. Set the rotary switch to the 
2. Press the MODE button to select the desired function. 
3. Connect  the  test  prods  of  the  leads  to  the  resistance  or  circuit  to  be 
tested. To avoid influence disconnect the resistant to be tested from the 
circuit. 
4. If the resistance is less than about 60 Ω, you hear an audible signal. If the 
circuit is open, the display shows "OL" or "1". 
 
Diode testing 
1. Set the rotary switch to the  
2. Press the MODE button to select the desired function. 
3. Touch the diode to be tested with the probes. The forward voltage shows 
400 to 700mV. The counter voltage shows „OL" or "1".   Defective devices 
show a value about 0 mV or „OL" or "1" in both polarities. 
 
 
 ‐ Position 
 ‐ Position 
 
 
EN 12 
 
 