ELECTROMAGNETIC COMPATIBILITY
GUIDANCE AND MANUFACTURE'S DECLARATION-ELECTROMAGNETIC IMMUNITY
Guidance and manufacture's declaration-electromagnetic immunity
The Pulse Oximeter is tended for use in the electromagnetic environment specified below. The user of
the Pulse Oximeter should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Power frequency
(50Hz) magnetic field
IEC 61000-4-8
FOR ALL EQUIPMENT AND SYSTEMS
IEC60601 test level
±6KV contact
±8KV air
3A/m
Compliance level
Electromagnetic environment-
guidance
±6KV contact
Floors should be wood,
±8KV air
concrete or ceramic tile. If floor
are covered with synthetic
material, the relative humidity
should be at least 30%.
Power frequency magnetic
3A/m
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment
ENGLISH • 25