SETTING THE BUZZER SUSPENDED PERIOD
An abnormality is indicated again by the buzzer sounding after a certain period has elapsed (the buzzer
recovery time) if the alarm mode continues even after the buzzer has already been silenced by pressing the
buzzer stop key (BUZZER). Always set the "buzzer suspended period" to avoid incorrect recognition of
alarm mode.
Setting range of buzzer suspended period:
between 10 and 60 minutes in 10-minute intervals (display of setting: between 010
and 060) or no recovery (display of setting; 000)
Initial setting (factory setting): 30 minutes (display of setting: 030)
The buzzer does not sound again when the setting is 000. The setting should be between 010 and
060 to ensure the safety of stored items.
For example, to change the setting of the buzzer suspended period from 30 minutes (initial setting) to 20
minutes
(The following is only an example. Choose the actual setting according to the desired buzzer suspended
period.)
Operation
1
Press the up arrow key for 5
2
seconds.
3 Press the up arrow key 5 times.
4 Press the scroll key once.
Press the up arrow key (as many
5
times as needed to reach the
desired figure).
6 Press the "Set" key once.
Press the up arrow key (as many
7
times as needed to reach the
desired figure).
8 Press the "Set" key once.
The unit will return from setting mode to temperature display mode automatically after 90 seconds if no
key is operated (auto-return function). In this case, the setting is not changed.
It is recommended that the buzzer suspended period is set when the freezer is not under alarm
conditions.
Key operated
The current chamber temperature is
----
displayed.
The current chamber temperature
changes to function code F00 and
the right-hand figure 0 blinks.
The display changes from F00 to F05.
The middle figure (0) starts to blink.
The display changes from F05 to F25.
The current setting (030) is displayed
SET
and the middle figure (3) blinks.
The display changes from 030 to 020.
The new setting is memorized and
SET
the current chamber temperature is
displayed.
Display after operation
25