REPLACEMENT OF WORN-OUT PARTS
Replacement of the battery for the power failure alarm
Replace the battery for the power-failure alarm every 3 years (when the battery indicator lights up) to
ensure the alarm will operate in the event of a power failure. Contact our sales representative or agent
to arrange the replacement of the battery when the battery indicator lights up.
The alarm function (blinking of alarm indicator, sounding of buzzer) will not operate when the battery for
the power-failure alarm is flat.
The alarm indicator blinks and the buzzer sounds as a result of power from the battery for the
power-failure alarm. The regular replacement of the battery for the power-failure alarm is important to
prevent an increase in chamber temperature in the event of unexpected circumstances.
<Important>
The used battery is a recyclable resource. Do not dispose of the battery. Always follow the procedure
for recycling.
Replacement of the fan motor
Replace the fan motor for the cooling circuit every 6 years (when the battery indicator blinks) to ensure
the appropriate operation of the freezer.
replacement of the fan motor for the cooling circuit when the battery indicator blinks.
The appropriate operation of the unit is maintained by cooling the compressor with a fan. The motor to
operate the fan (the fan motor for the cooling circuit) has a limited lifespan. A fan motor with diminished
performance may lead to reduced cooling performance.
CALIBRATION
During continuous operation, the following service tasks must be performed:
・ Perform a temperature calibration at least once a year.
For temperature calibration, contact our sales representative or agent.
32
Contact our sales representative or agent to arrange the