Sartorius YDO01C-232 Instrucciones De Instalación página 6

Salida de datos para interfaz unicom de combics
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

1
3
Configuration of the solder bridges
Current interface, 4 to 20 mA
1
Current interface, 0 to 20 mA
1
Voltage interface, 0 to 5 V
1
Operation with internal voltage supply
3
4
Operation with external voltage supply
3
4
Operation as an electrically isolated
interface
3
4
6
The factory defaults are as follows (see the illustration on the left):
– Operation as a current interface
– Range of current intensity: 4 to 20 mA
– Voltage supply: internal
2
4
Operation as a Current Interface in the 4 to 20 mA Range (Factory Setting):
§ Solder bridges 1 and 2 must be configured as shown in the illustration on the left.
2
Operation as a Current Interface in the 0 to 20 mA Range:
§ Re-solder bridge 1 as shown in the illustration on the left.
2
§ Solder bridge 2 must be configured as shown in the illustration on the left. This
corresponds to the factory default configuration.
Operation as a Voltage Interface (0 to 5 V):
§ Open solder bridge 1 so that the setting corresponds to that shown here on the left.
2
§ Re-solder bridge 2 as shown in the illustration.
Operation with Internal Voltage Supply (Factory Setting):
§ Solder bridges 3 and 4 must be configured as shown in the illustration on the left.
Operation with External Voltage Supply:
§ Open solder bridge 3.
§ Re-solder bridge 4 as shown in the illustration.
Operation as an Electrically Isolated Interface:
§ Configure solder bridges 3 and 4 for "external supply." To do this, open solder bridge
3 and re-solder bridge 4 as shown in the illustration.
§ If the interface is powered by an external AC adapter, it must be electrically isolated.
Be sure to comply with the requirements of CE conformity.
Notes
– The shielding in the current interface connecting cable is connected at one end to the
housing of the indicator.
– The indicator is connected to the protective grounding conductor.
– The current interface cannot be operated when the indicator is powered by the external
rechargeable battery pack.
§ The YDA01C-20MA interface module (Option A4) is installed directly on the digital PCB
in the Combics indicator (models CISL., CIS., CW.P, CW.S). For details, please see
"Installation in the Indicator."

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ydo01c-485Yda01c-20maYdo01c-dpYdo01c-en

Tabla de contenido