9. APPENDIX
9.1. ERROR CODES
During initial test runs or during normal operation, the scale may display an error message. The meaning of these
error messages is described below.
If an error message appears, repeat the procedure that produced the message: turning on the scale, calibration, or
any other function. If the error message still appears you should contact your dealer for technical service.
ERROR CODE
ERR 4
ERR 5
ERR 6
To view the A/D converter divisions, press key
ming the initial check at power-on. Press
9.2 LIST OF WEIGHT UNITS
Nº
UNIT SYMBOL
1
2
3
4
5
6
7
8
9
10
11
12
13
14
DESCRIPTION
The leading zero is greater than allowed (4% of the
maximum capacity) when the scale is turned on or
when the key is pressed
Keyboard error.
The A/D converter divisions are not correct when
the scale is turned on.
g
g
gram (high resolution)
ct
lb
oz
d
gn
ozt
dwt
mm
tl.T
tl.c
Tl.t
t
POSSIBLE CAUSES
There is weight in the tray when the scale is turned
on.
.
Excessive weight in the tray when zeroing the scale.
Incorrect calibration of the scale.
Damaged loading cell.
Damaged electronic components.
Incorrect operation of the scale.
The platform is not installed.
Damaged loading cell.
Damaged electronic components.
and key
at the same time while the scale is perfor-
to return to normal function.
UNIT NAME
gram
net carat
= 0,2 g
pound
= 453,59237 g
ounce
= 28, 349523000 g
drachma
= 1,7718450 g
grain
= 0,06479900 g
troy ounce
= 31,1034770 g
pennyweight
= 1,55517400 g
momme
= 3,749996 g
tael twn
= 37,49996 g
tael chn
= 37,799375 g
tael troy
= 37,4290020 g
tola
= 11,6638040 g
USER'S GUIDE
CONVERSION
EN
EN
33