P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may need
to perform preventive maintenance checks more often based on your level of product usage.
Remove product from service before performing preventive maintenance.
N N o o t t e e - - Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items for the standard models:
Zipper and covers (top and bottom) are free of tears, cuts, holes, or other openings
Internal components for signs of staining from fluid ingress or contamination by fully unzipping the covers
Labels for legibility, proper adherence, and integrity
Handles are free of rips or cracks and stitching is intact
Foam has not degraded or come apart
Fire barrier for rips, cracks, or other visible signs of damage
Inspect the following items for the behavioral health models:
Covers (top and bottom) are free of tears, cuts, holes, or other openings
Labels for legibility, proper adherence, and integrity
Product serial number:
Completed by:
Date:
C C a a r r e e a a n n d d m m a a i i n n t t e e n n a a n n c c e e
W W A A R R N N I I N N G G
• Do not wash the internal components of this support surface. Discard the support surface if a contamination is found
inside.
• Do not immerse the product in cleaning or disinfectant solutions.
• Do not allow liquid to pool on the product.
• Always inspect support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned zippers
before each use. If compromised immediately remove the support surface from service.
• Always make sure that you wipe each product with clean water and dry each product after you clean. Some cleaning
agents are corrosive in nature and may cause damage to the product. Failure to follow these cleaning instructions may
void your warranty.
C C A A U U T T I I O O N N
• Do not allow liquid to seep into the zipper area or watershed cover barrier when you clean the underside of the support
surface. Fluids allowed to come in contact with the zipper may leak into the support surface core.
• Do not iron, dry-clean, or tumble dry the support surface covers.
• Do not power wash the support surface as this may damage the product.
• Always dry the support surface covers before you store, add linens, or place a patient on the surface. A dry product
helps to prevent impaired performance.
• Do not over expose the covers to higher concentration chemical solutions as these may degrade the covers.
• Do not use accelerated hydrogen peroxides or quaternaries that contain glcyol ether >3% as these chemicals may
damage the support surface cover.
• Failure to follow manufacturing instructions may also affect useful life of the support surface cover.
EN
10
2815-009-001 Rev E.0