S4 PIONEER X-ray Spectrometer
Introduction
Servicing Precautions
To ensure that the spectrometry system will operate in a reliable and precise way, the maintenance
intervals must be obeyed as recommended by Bruker AXS.
All repairs, adjustments and alignments performed on any components of the spectrometry system
must be carried out only by authorized personnel.
WARNING
When the equipment is connected to the mains supply, some terminals of the mains distribution unit
may be live. Therefore it is absolutely necessary to switch off the external mains supply before open-
ing the side panels. It is not sufficient to just press the S4's 'Power OFF' button. The mains supply
must be switched off externally on the customer's side (wall socket or external switch) (see chapter
'Control Electronics-Mains Distribution Board' in this manual).
Cleaning the Spectrometry System
For cleaning the side panels and the interior of the sample magazine use dry cleaning utensils only.
Do not use water or aggressive cleansing agents. The openable transparent sample magazine cover
can be cleaned with any standard household-type window detergent.
CAUTION
Before starting cleaning you must turn off all spectrometry system power.
DOC-M80-ZXX004 V6 – 03.2007
2-13