7) Inspect the apical flutes of the ProTaper Next
finished, the correspondingly sized gutta percha master cone or size verifier may be fitted, and the canal
is ready for disinfection.
8) Alternatively, gauge the foramen with a size 025 hand file and, if this file is snug at length, the canal is
shaped and ready for disinfection.
9) If the size 025 hand file is loose at length, then continue shaping with the ProTaper Next
when necessary, the ProTaper Next
instrument with the 030, 040 or 050 hand files, respectively.
During protocol of use, irrigate, recapitulate with a small-sized hand file after each sequential
ProTaper Next
instrument, then re-irrigate.
®
B EN PTN0 DFU WEB / Rev.05 / 05-2017 (Old ZF 1902123.EN)
X2 file; if they are loaded with dentin, then the shape is
®
X4 (040/06) or ProTaper Next
®
®
X5 (050/06), gauging after each
®
X3 (30/07) and,
3/4