Getting Started - Minebea Intec Combics CAIXS2 Advertencias De Instalación Y De Seguridad

Visor con seguridad intrínseca para el uso en zonas con peligro de explosión
Ocultar thumbs Ver también para Combics CAIXS2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50

Getting Started

6
Installationsanleitung CAIXS2
Getting Started
1.) Connect weighing platform to the indicator.
2.) CAIXS2 Indicators have an intrinsically safe data interface which can be
connected to a computer (or any other peripheral device) using a barrier
(e.g. YDI05-Z).
3.) Connect the AC adapter.
4.) Configure the analog/digital converter (ADC).
5.) Carry out an alignment: Adjustment.
Connecting Weighing Platforms
You can connect any intrinsically safe, analog Minebea Intec platform to your
CAIXS2 Indicator. Refer also to the Verification of Intrinsic Safety, the EC Type
Examination Certificate for the CAIXS2 and the load cell or analog weighing
platform to be connected.
The load cell should be connected by a certified technician who has received
3
specialized training from Minebea Intec. Any installation work that does not
conform to the instructions in this manual results in forfeiture of all claims
under the manufacturer's warranty.
Make sure the CAIXS2 is disconnected from the power before starting any
3
connection work.
Set up the weighing platform (see operating instructions for the weighing
t
platform).
Place the cable from the weighing platform next to the indicator.
t
Open the Combics indicator:
t
Loosen the ten cap nuts on the front panel. Remove the front panel.
Installing Connection and Interface Cables
The cable gland (IP69K protection) is pre-mounted on the indicator.
3
Please use extreme caution when performing any work on the equipment that
affects this cable gland.
You must use a torque wrench. The torque for this cable gland is 5 Nm.
Preparing the Cable
Strip approx. 14 cm from the end of the cable.
t
Shorten the shielding to approx. 2 cm and pull back over the insulation.
t
Strip approximately 5 mm of the insulation from the wires of the connecting
t
cable and affix ferrules to the wire ends.

Publicidad

Tabla de contenido
loading

Tabla de contenido