MANUAL DE INSTRUCCIONES CODIGO 80222 REV A
J
.P. SELECTA s.a.
NOV-2007
9 Maintenance
Before of remove the cover to handle inside, unplug the the equipment
from the socket.
The handle of the electronic circuit boards by unauthorized personnel
may cause damages. Make sure to giv e the equipment to an
authorised technical assistance.
9.1 CLEANING
We recomend to use a cotton cloth wet with a water-alcohol disolution.
Inner cleaning and decontamination of sampler or products run out
into the centrifuge vessel
The user has the responsability to do the decontamination according to
each dangerous product that could be run out inside the centrifuge vessel.
In case of samples splashed in the vessel clean it with a biological disinfectant.
9.2 MAINTENENCE OPERATION
The followi ng parts make the user protectiun up againts impacts of
unexpected parts thrown so a quarterly inspection of them are required:
• Lid, hinges, locks and enclosures: Inspect visually possible craking
formations, deformations, rust stains, etc which might sign a possible
dropping material resistance.
• Rubber legs: Check that the rubber legs keep the flexibility and grip on
the working surface.
9.3 SPARE PARTS
The spare parts more commons:
Spare description
Electromagnetic lock
Main switch
Motor
Tachometer sensor circuit board
Control panel circuit board
Power circuit board
Mains filter
Transformar 115-115/18V 1.2A
10 Guarentee
This product is guaranteed for one year. The guarantee does not cover
damage ca used by incorrec t use or causes b eyond the control of J.P.
SEL ECTA,S.A.
Any manipulation of the apparatus by personnel not authorized by J.P.
SELECTA,S.A. automatically cancels the guarantee.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.jpselecta.es
(Sujetas a modificaciones sin previo aviso)
Tel 34 937 700 877 Fax 34 937 702 362