Renishaw UCC T3 Guia De Instalacion página 32

Tabla de contenido

Publicidad

UCC T3 installation guide
www.renishaw.com
Bypass plug
E-STOP safety test
The E-STOP buttons should be regularly inspected to ensure that they are in good working order and easily accessible. It should also be
tested as follows:
1. Ensure that the probe head is free to move in all six directions, using the MCU joystick.
2. Hit the emergency stop button. For some systems, the button must then be unlocked by twisting it, as indicated by the arrows on the
button.
3. Ensure that the probe head cannot be moved, using the MCU joystick.
4. Re-engage the servos using the button on the MCU and check that the probe head can now move again.
Issued 04 2019
32